Radiographic measurement of the congruence angle according to Merchant: validity, reproducibility, and limits

Abstract

Purpose

The objective of this study was to analyze the intra- and interobserver variability of this measurement according to a strict methodology and on a representative sample of the general population, as well as to identify the possible difficulties of measurement in case of patellar or trochlear dysplasia.

Methods

This observational study involved radiographic analysis by three independent observers of a total of 50 patients who had a loaded patellofemoral X-ray taken with the knee flexed to 45°. An initial reading was taken to measure the angle of the trochlear sulcus, the Merchant angle, and to classify the knees according to a possible trochlear dysplasia and/or patellar dysplasia according to Wiberg. A second measurement was then performed to analyze intraobserver agreement. Interobserver agreement was measured on all radiographic measurements (n = 100).

Results

The Merchant patellofemoral congruence angle showed good intraobserver concordance ranging from 0.925 (95% CI 0.868–0.957) to 0.942 (95% CI 0.898–0.967), as well as interobserver concordance ranging from 0.795 (95% CI 0.695–0.862) to 0.914 (95% CI 0.872–0.942). Poor results were found in terms of interobserver concordance on the measurement of the Merchant angle in case of stage 3 Wiberg patella ranging from 0.282 (95% CI −0.920 to 0.731) to 0.611 (95% CI 0.226–0.892).

Conclusion

Congruence angle is one of most commonly used measurements for patellar tracking. However, the convexity of the patellar surface makes it difficult to identify the patellar apex on its intraarticular facet, making the measurement of the Merchant congruence angle unreliable and not very reproducible in cases of stage 3 Wiberg patella.

Registration N°IRB 2021/139

Introduction

Assessment of patellofemoral congruence is performed in routine practice in cases of patellofemoral instability [1, 2], painful patellar syndrome [3], or patellofemoral osteoarthritis [4]. In these cases, the analysis of patellofemoral congruence makes it possible to justify a surgical treatment choice in case of chronic instability or chronic patellofemoral pain (tibial tuberosity osteotomy, external patellar release, axial correction and/or derotation osteotomy, medial patellofemoral ligament reconstruction) [5,6,7,8]. In the literature, the Merchant angle or congruence angle is currently the most widely used radiographic measure for assessing this patellofemoral congruence [9,10,11]. However in 2013, a meta-analysis demonstrated that there was insufficient evidence to determine the reliability, validity, sensitivity, and specificity of this congruence angle [12]. In fact, the various studies found were conducted either on a population with healthy knees, representative of only 20% of the general population, or on series of cases of patellofemoral instability, with great variability in the radiographic acquisition protocols [2]. Moreover, no intraobserver or intraobserver analysis was performed of this radiographic measurement according to the level of trochlear or patellar dysplasia.

The main objective of this study was to analyze the intra- and interobserver variability of this radiographic angle of patellofemoral congruence described by Merchant according to a strict methodology and on a sample representative of the general population. The second objective was to analyze the variations of this angular measurement according to a possible patellar or trochlear dysplasia.

Materials and methods

Study population

This observational study allowed the radiographic analysis of a total of 50 patients who performed a loaded patellofemoral X-ray with the knee flexed to 45° according to the service protocol. These 50 radiological images were selected in chronological order of acquisition as of 1 January 2021, if all inclusion criteria were met. The inclusion criteria were: patient with closed physis and having performed a 45° patellofemoral X-ray under load on the native joint, and a profile X-ray to confirm the degree of flexion and to look for possible trochlear dysplasia. Patients with prosthetic or osteosynthesis material (n = 12), fractured patients (n = 4), and those who refused to participate in the study or who were under legal protection (n = 2) were excluded.

Measurement method

In 1974, Merchant et al. [9] proposed a method of assessing the degree of patellofemoral congruence by radiographic measurement of the patellofemoral congruence angle. To this day, this angle remains the reference radiographic method for measuring this congruence. To make this measurement (Fig. 1), the bisector of the sulcus angle is drawn to establish a zero reference line. A second line is then projected from the apex of the sulcus angle to the lowest point of the apex of the intraarticular facet of the patella. The angle measured between these two lines is the congruence angle described by Merchant [9]. If the apex of the patellar joint ridge is lateral to the zero line, then the congruence angle is positive. If it is medial, then the angle is negative (Figs. 2, 3).

Fig. 1
figure 1

Method of measuring the patellofemoral congruence angle on a patellofemoral X-ray at 45° of flexion. Draw the angle of the sulcus (BAC) from the apex of the medial (B) and lateral (C) condyle, and identify the deepest point of the intercondylar sulcus (A). Draw the line AD through the apex of the sulcus angle (A) and the apex of the articular patellar ridge (D). Merchant angle is therefore the angle between the line DA and the bisector of the sulcus angle [9]

Fig. 2
figure 2

Example of measurement of the patellofemoral congruence angle according to Merchant on a patellofemoral image of the knee flexed at 45° ((A) native knee, (B) same knee after correction by distal femoral osteotomy)

Fig. 3
figure 3

Radiograph of a Wiberg stage 3 patella case with intra- and interobserver variability in measurement

Assessment criteria

Radiographic analysis was performed independently by three senior orthopedic surgeons (J.M., T.B., B.S.) of which only one is specialized in the management of the patellofemoral joint (J.M.). All patellofemoral X-rays were analyzed. After anonymization of the radiographs, a first reading allowed each observer to measure the angle of the trochlear sulcus as well as the Merchant angle. A second measurement, still anonymized and in disorder, was performed to analyze the intraobserver concordance of the measurement of the Merchant congruence angle between all the participants taken in pairs. Interobserver concordance was measured on all radiographic measurements, i.e., a total of 100 radiographs. To analyze the intra- and interobserver variability according to a possible patellar or trochlear dysplasia that could complicate the measurements, an analysis of the intra- and interobserver concordances of the Merchant angle in subgroups was performed. The degree of trochlear dysplasia according to Dejour as well as the classification of the patella according to Wiberg were analyzed after the first measurements during a collegial meeting that allowed a consensus to be reached between the three observers on all the X-rays.

In case of poor agreement on the measurement of Merchant angle, a comparison was made with the measurement of the trochlear sulcus. This made it possible to identify the source of confusion on the radiographic measurement between the positioning of the marker on the apex of the patella or the measurement of the angle of the sulcus by the positioning of the markers on the tops of the condyles in coronal section.

Statistical analysis

Data were collected in an Excel spreadsheet (Microsoft, Richmond, WA, USA) and analyzed with IBM SPSS software release 21 (IBM Corporation, Armonk, NY, USA) via a protocol validated by the institutional review board (IRB) dependent on the research department of our institution (IRB reference no. 2021/139). Continuous variables were described by their mean and standard deviation; and categorical variables by their number and percentage.

Intra- and interobserver agreement of Merchant angle measurement as well as trochlear sulcus angle was measured by the intraclass correlation coefficient, with its 95% confidence interval. In accordance with Nunnally and Berstein [13], an intraclass correlation coefficient greater than 0.7 could be considered good.

Results

Table 1 presents the distribution of the different patella dysplasia according to the Wiberg classification as well as possible trochlear dysplasia among the 100 radiographic cases studied.

Table 1 Distribution of patients according to the Wiberg and Dejour classification (N = 50)

Intra- and interobserver correlations of the Merchant angle, the degree of trochlear dysplasia, and patellar dysplasia are presented in Tables 2, 3, and 4.

Table 2 Intra-class correlation coefficients for Merchant congruence angle measures
Table 3 Results of the measurement of intra- and interobserver concordance according to the degree of trochlear dysplasia according to Dejour
Table 4 Results of the measurement of intra- and interobserver concordance based on the Wiberg classification of the patella

In all the patients included, the intraobserver concordance of the Merchant angle was very good, ranging from 0.926 (95% CI 0.869–0.956) to 0.943 (95% CI 0.892–0.965). They were also good for interobserver concordance, ranging from 0.797 (CI 95% 0.690–0.865) to 0.912 (CI 95% 0.870–0.945).

All these results remained good with and without trochlear dysplasia (Table 3).

Regarding the Wiberg classification, poor results were found in terms of interobserver concordance on the measurement of the Merchant angle in case of Wiberg patella stage 3 ranging from 0.285 (95% CI −0.918 to 0.738) to 0.608 (95% CI 0.222–0.897).

Table 5 presents the results of the intraclass correlation of the angular measurements of the sulcus in case of stage 3 Wiberg patella. Since the results of this angular measurement are very satisfactory (ICCC of 0.922–0.941), the positioning of the landmark on the apex of the patella seems to be the cause of the poor results observed on the intraclass correlation measurements in case of stage 3 Wiberg patella.

Table 5 Evaluation of trochlea sulcus measurement by measuring the intraclass correlation coefficient

Discussion

This study showed very good results in terms of intra- and interobserver concordance regarding the radiographic measurement of patellofemoral congruence angle. This corresponds to the results observed in the literature on the validity and reproducibility of this measurement in a population of patients with patellar instability, with better results by radiographic evaluation method than by computed tomography (CT) scan or MRI measurement [12].

Although it is a useful two-dimensional radiographic measurement to characterize the geometry of the patellofemoral joint, some authors feel that it provides limited information regarding the complex interface of the contact surfaces or the transmission of load through the patellofemoral joint compared with CT or MRI measurements. Recently, Lee et al. [14] established threshold values for patellofemoral parameters after three-dimensional CT reconstruction, including the congruence angle. Although these values facilitate the diagnosis and treatment planning of patellofemoral disorders in skeletally mature patients, their acquisition modalities remain complex, and were performed in the supine position with 15° of knee flexion. Regarding the evaluation and measurement by MRI, Ye et al. [15] found very poor interobserver reliability regarding the measurement of this congruence angle (ICCs 0.325–0.380), in contrast to the other measurements of the patellofemoral joint: Fulkerson angle [16], Laurin angle [17], patellar tilt angle (PTA), lateral patellar displacement (LPD), and bisect offset ratio (ICCs > 0.8). For the authors, these poor results could be explained by the changes in knee position during iconographic acquisition and/or the difficulty in identifying the patellar apex [15].

This last finding can be correlated with the results of our secondary objective. Poor results in terms of intra- and interobserver concordance were found in cases of stage 3 Wiberg patella, whereas the sulcus measurement obtained an intraclass correlation coefficient greater than 0.7. The convexity of the patella facet joint makes radiographic measurement of the Merchant congruence angle difficult and not reproducible. However, patients with Wiberg stage 3 patellar joints appear to be at greater risk of patellofemoral osteoarthritis and therefore require a radiographic evaluation [18].

While the reference patellofemoral X-ray protocol at 45° of flexion described by Merchant should be performed in the supine position, imaging protocols tend to perform these same acquisitions in the weight-bearing position with the same reliability [19]. In this work, the authors wanted to be in loaded conditions. Indeed, the movement of the patella in its patellofemoral joint is affected by the contraction of the quadriceps muscle [20] as well as the joint load [21, 22]. Compared with the supine position protocol, the 45° loaded patellofemoral X-ray would allow for a decrease in the congruence angle [23]. Consideration of both loaded and unloaded images appears to provide additional information in the radiographic evaluation of the patellofemoral joint.

This study has a number of limitations related primarily to its retrospective nature. Our study did not allow us to ensure the correct positioning of the knee at 45° during the various acquisitions. In their systematic review, Nord et al. [24] observed great heterogeneity in imaging protocols and poor reproducibility of axial patellofemoral X-rays. Although this was not detrimental to our results, this variability could lead to a disparity in the results of radiographic images used for clinical decision making [24].

In our study, all three observers were senior orthopedic surgeons but only one was experienced in patellofemoral radiological analysis. For a large number of authors [25, 26], the level of experience seems to play an important role in the evaluation and measurement of patellofemoral parameters with a significant learning curve. As an example, Smith et al. [25] found intraobserver variations in congruence angle measurement that could range from 0.8° to 18.4° for the same imaging. These variations tended to balance out according to the observer’s level of experience. Although the results were similar between the “naïve” and the “expert” observers, This notion should be taken into consideration when analyzing the results of this work which were initially measured by three senior orthopedic surgeons.

Conclusion

Congruence angle is one of most commonly used measurements for patellar tracking. The patellofemoral congruence angle according to Merchant shows good intra- and interobserver concordance in a population with and without patellofemoral dysplasia, apart from cases of Wiberg stage 3 patella. Indeed, the convexity of the patellar surface makes it difficult to identify the patellar apex on its intraarticular facet, making the measurement of this angle unreliable and not very reproducible in this case.

Availability of data and materials

Data will not be shared.

References

  1. McCrum E, Cooper K, Wittstein J, French RJ (2021) Imaging of patellofemoral instability. Clin Sports Med 40(4):693–712

    Article  Google Scholar 

  2. White AE, Otlans PT, Horan DP, Calem DB, Emper WD, Freedman KB et al (2021) Radiologic measurements in the assessment of patellar instability: a systematic review and meta-analysis. Orthop J Sports Med 9(5):2325967121993179

    Google Scholar 

  3. Drew BT, Redmond AC, Smith TO, Penny F, Conaghan PG (2016) Which patellofemoral joint imaging features are associated with patellofemoral pain? Systematic review and meta-analysis. Osteoarthritis Cartilage 24(2):224–236

    Article  CAS  Google Scholar 

  4. Mehl J, Feucht MJ, Bode G, Dovi-Akue D, Südkamp NP, Niemeyer P (2016) Association between patellar cartilage defects and patellofemoral geometry: a matched-pair MRI comparison of patients with and without isolated patellar cartilage defects. Knee Surg Sports Traumatol Arthrosc 24(3):838–846

    Article  Google Scholar 

  5. Swarup I, Elattar O, Rozbruch SR (2017) Patellar instability treated with distal femoral osteotomy. Knee 24(3):608–614

    Article  Google Scholar 

  6. Barnavon T, Odri GA, Vendeuvre T, Labrada-Blanco O, Bordes M, Renard G et al (2020) Medial closing-wedge distal femoral varus osteotomy: symptoms and functional impact in cases of associated patellofemoral osteoarthritis. A two-year follow-up prospective pilot study. Knee 27(3):615–623

    Article  CAS  Google Scholar 

  7. Zhang H, Ye M, Liang Q (2020) Clinical outcomes after medial patellofemoral ligament reconstruction with suture fixation of the gracilis tendon via transosseous tunnels. Orthop J Sports Med 8(2):2325967119900373

    Google Scholar 

  8. Cregar WM, Huddleston HP, Wong SE, Farr J, Yanke AB (2022) Inconsistencies in reporting risk factors for medial patellofemoral ligament reconstruction failure: a systematic review. Am J Sports Med 50(3):867–877

    Article  Google Scholar 

  9. Merchant AC, Mercer RL, Jacobsen RH, Cool CR (1974) Roentgenographic analysis of patellofemoral congruence. J Bone Joint Surg Am 56(7):1391–1396

    Article  CAS  Google Scholar 

  10. Merchant AC (1988) Classification of patellofemoral disorders. Arthroscopy 4(4):235–240

    Article  CAS  Google Scholar 

  11. Merchant AC (2001) Patellofemoral imaging. Clin Orthop Relat Res 389:15–21

    Article  Google Scholar 

  12. Smith TO, Davies L, Toms AP, Hing CB, Donell ST (2011) The reliability and validity of radiological assessment for patellar instability. A systematic review and meta-analysis. Skeletal Radiol 40(4):399–414

    Article  Google Scholar 

  13. Nunnally JC, Bernstein IH. Psychometric Theory. In: 3rd ed. New York: MsGraw-Hill; 1994.

  14. Lee KW, Seo D-K, Bae J-Y, Ra HJ, Choi S-J, Kim JK (2021) Usefulness of three-dimensional computed tomography for patellofemoral measurement. Knee Surg Sports Traumatol Arthrosc 30:1423

    Article  Google Scholar 

  15. Ye Q, Yu T, Wu Y, Ding X, Gong X (2019) Patellar instability: the reliability of magnetic resonance imaging measurement parameters. BMC Musculoskelet Disord 20(1):317

    Article  Google Scholar 

  16. Schutzer SF, Ramsby GR, Fulkerson JP (1986) The evaluation of patellofemoral pain using computerized tomography. A preliminary study. Clin Orthop Relat Res 204:286–293

    Article  Google Scholar 

  17. Laurin CA, Lévesque HP, Dussault R, Labelle H, Peides JP (1978) The abnormal lateral patellofemoral angle: a diagnostic roentgenographic sign of recurrent patellar subluxation. J Bone Joint Surg Am 60(1):55–60

    Article  CAS  Google Scholar 

  18. Dai Y, Yin H, Xu C, Zhang H, Guo A, Diao N (2021) Association of patellofemoral morphology and alignment with the radiographic severity of patellofemoral osteoarthritis. J Orthop Surg Res 16(1):548

    Article  Google Scholar 

  19. Gharaibeh MA, Monk E, Chen DB, MacDessi SJ (2018) Evaluation of the patellofemoral joint in total knee arthroplasty: validation of the weight bearing merchant radiographic view. Knee 25(6):1262–1271

    Article  Google Scholar 

  20. Brossmann J, Muhle C, Schröder C, Melchert UH, Büll CC, Spielmann RP et al (1993) Patellar tracking patterns during active and passive knee extension: evaluation with motion-triggered cine MR imaging. Radiology 187(1):205–212

    Article  CAS  Google Scholar 

  21. Powers CM, Ward SR, Fredericson M, Guillet M, Shellock FG (2003) Patellofemoral kinematics during weight-bearing and non-weight-bearing knee extension in persons with lateral subluxation of the patella: a preliminary study. J Orthop Sports Phys Ther 33(11):677–685

    Article  Google Scholar 

  22. McWalter EJ, Hunter DJ, Wilson DR (2010) The effect of load magnitude on three-dimensional patellar kinematics in vivo. J Biomech 43(10):1890–1897

    Article  CAS  Google Scholar 

  23. Kim T-H, Lee J-S, Oh K-J (2014) Discrepancies of patellofemoral indices between supine and standing merchant views. Knee Surg Relat Res 26(1):20–26

    Article  Google Scholar 

  24. Nord A, Agel J, Arendt EA (2014) Axial knee radiographs: consistency across clinic sites. Knee Surg Sports Traumatol Arthrosc 22(10):2401–2407

    Article  Google Scholar 

  25. Smith TO, Cogan A, Patel S, Shakokani M, Toms AP, Donell ST (2013) The intra- and inter-rater reliability of X-ray radiological measurements for patellar instability. Knee 20(2):133–138

    Article  CAS  Google Scholar 

  26. Nicolaas L, Tigchelaar S, Koëter S (2011) Patellofemoral evaluation with magnetic resonance imaging in 51 knees of asymptomatic subjects. Knee Surg Sports Traumatol Arthrosc 19(10):1735–1739

    Article  CAS  Google Scholar 

Download references

Acknowledgements

Many thanks to Professor A.C. Merchant of Stanford University (Los Altos, CA 94022, USA) for his precious cognitive aids, his kindness, and his availability.

Funding

No funding source.

Author information

Authors and Affiliations

Authors

Contributions

S.B., B.T., M.J.: data collection; D.M.: statistical analysis; V.T., M.S.: redaction and data collection. All authors read and approved the final manuscript.

Corresponding author

Correspondence to M. Severyns.

Ethics declarations

Ethics approval and consent to participate

IRB approval N°IRB 2021/139.

Consent for publication

Yes (N°IRB 2021/139).

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Severyns, M., Mallet, J., Santoni, B. et al. Radiographic measurement of the congruence angle according to Merchant: validity, reproducibility, and limits. Knee Surg & Relat Res 35, 1 (2023). https://doi.org/10.1186/s43019-023-00175-5

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s43019-023-00175-5

Keywords