The use of imageless navigation to quantify cutting error in total knee arthroplasty

Abstract

Purpose

Navigated total knee arthroplasty (TKA) improves implant alignment by providing feedback on resection parameters based on femoral and tibial cutting guide positions. However, saw blade thickness, deflection, and cutting guide motion may lead to final bone cuts differing from planned resections, potentially contributing to suboptimal component alignment. We used an imageless navigation device to intraoperatively quantify the magnitude of error between planned and actual resections, hypothesizing final bone cuts will differ from planned alignment.

Materials and methods

A retrospective study including 60 consecutive patients undergoing primary TKA using a novel imageless navigation device was conducted. Device measurements of resection parameters were obtained via attachment of optical trackers to femoral and tibial cutting guides prior to resection. Following resection, optical trackers were placed directly on the bone cut surface and measurements were recorded. Cutting guide and bone resection measurements of both femoral and tibial varus/valgus, femoral flexion, tibial slope angles, and both femoral and tibial medial and lateral resection depths were compared using a Student's t-test.

Results

Femoral cutting guide position differed from the actual cut by an average 0.6 ± 0.5° (p = 0.85) in the varus/valgus angle and 1.0 ± 1.0° (p = 0.003) in the flexion/extension angle. The difference between planned and actual cut measurements for medial and lateral femoral resection depth was 1.1 ± 1.1 mm (p = 0.32) and 1.2 ± 1.0 mm (p = 0.067), respectively. Planned cut measurements based on tibial guide position differed from the actual cut by an average of 0.9 ± 0.8° (p = 0.63) in the varus/valgus angle and 1.1 ± 1.0° (p = 0.95) in slope angle. Measurement of medial and lateral tibial resection depth differed by an average of 0.1 ± 1.8 mm (p = 0.78) and 0.2 ± 2.1 mm (p = 0.85), respectively.

Conclusions

Significant discrepancies between planned and actual femoral bone resection were demonstrated for flexion/extension angle, likely the result of cutting error. Our data highlights the importance of cut verification postresection to confirm planned resections are achieved, and suggests imageless navigation may be a source of feedback that would allow surgeons to intraoperatively adjust resections to achieve optimal implant alignment.

Introduction

Total knee arthroplasty (TKA) is one of the most common and effective surgical interventions for end-stage osteoarthritis of the knee joint [1, 2]. While advances in orthopedic technology have improved postsurgical outcomes, implant malalignment remains a predominate issue, with unsatisfactory prosthesis positioning reported in up to 30% of primary TKA procedures [3,4,5,6]. The well-documented association between malalignment, poor patient outcomes, and lower rates of implant survivability underscores the importance of optimizing component alignment [7,8,9]. Consequently, efforts to lessen both the prevalence and magnitude of alignment errors help mitigate associated complications, increasing the likelihood of long-term postoperative success and favorable patient outcomes [4, 10].

While the causes of malalignment are multifactorial, cutting errors associated with femoral and tibial resection are known to play a significant role [11, 12]. Bone cuts typically rely on an oscillating bone saw, the blade of which is subject to deflection, which may lead to movement of the cutting guide. This deflection has been shown to contribute to deviations from the planned plane of resection and, ultimately, to errors in component placement [11,12,13]. Cutting errors have been reported to cause deviations in component position of up to 4° in the coronal plane [11, 14] and 10° in the sagittal plane [11, 15], a magnitude of error that exceeds the ideal clinical range of ± 3° from the mechanical axis of the joint and can contribute to limb malalignment [9,10,11, 16,17,18]. Malalignment leads to uneven load distribution [6, 19] and suboptimal restoration of joint biomechanics [4, 5], both of which can adversely affect long-term procedural success, significantly elevate the rate of prosthesis failure, and increase the likelihood of future revision surgery [3, 5, 9, 20]. Furthermore, resection errors can result in uneven bone surfaces, causing gaps between bone and implant that have deleterious effects on alignment, implant adhesion in cemented cases, and bone ingrowth in uncemented implants [11, 12, 21]. Given the adverse impact of resection error on alignment, a better understanding of the magnitude and prevalence of this error is critical to mitigating it.

The integration of computer-assisted navigation systems (CAS) in TKA procedures has grown in popularity due to their ability to improve accuracy of component placement by providing real-time feedback on resection parameters based on tibial and femoral cutting guides [22]. This helps facilitate alignment to ensure optimal cut parameters are achieved, leading to an improvement in both component and limb alignment compared with conventional methods [23]. Additionally, CAS offers the ability to verify bone cuts intraoperatively postresection, allowing surgeons to make adjustments and modifications prior to implant placement that may optimize alignment. We utilized these features in a novel CAS system to compare planned and actual cut measurements for the femur and tibia in a cohort of patients undergoing primary TKA with computer navigation assistance. The difference between the planned and actual cuts provides an indication of the error associated with bone resection during primary TKA procedures. We hypothesized that final bone cuts will minimally differ from planned alignment.

Materials and methods

Study design and patient population

We conducted a retrospective review of patients who underwent a primary TKA with the assistance of imageless computer navigation between October 2019 and December 2019 at a single medical center. All procedures were performed by one of three senior fellowship-trained and board-certified surgeons specialized in total joint arthroplasty. Inclusion criteria required successful positioning of femoral and/or tibial cutting guides using imageless navigation and intraoperative measurement of pre- and postresection cuts by the imageless computer navigation device. Procedures in which the navigation device was removed prior to the recording of key measurements, or the measurements were not recorded or obtained from the device were excluded from analysis. This study received Institutional Review Board (IRB) approval from the institutional ethics board prior to data collection.

Surgical technique

All procedures were performed via the medial parapatellar approach and used imageless computer navigation (Intellijoint KNEE™, Intellijoint Surgical, Kitchener, ON). Given this navigation device accommodates either a femur- or tibia-first workflow, the TKA procedures included in this study represent both workflows. The device is composed of a camera, optical trackers, and a computer workstation that sits outside of the sterile field and is controlled by the surgeon via buttons on the camera (Fig. 1). The workflow for tibial cuts is as follows: after the primary incision and exposure of the femur and tibia, a bone screw is drilled into the tibia and the optical bone tracker is attached. The bone screw may be inserted either on the articular or extraarticular surface of the proximal tibia; however, for cut verification, extraarticular placement of the screw is required. The tibia is registered using the optical probe tracker by defining the medial and lateral malleoli, the tibial center, the anteroposterior (AP) axis, and the medial and lateral tibial plateaus. The optical probe tracker is then attached to the cutting guide, providing real-time feedback of the prospective cut measurements. Cutting guide position can be adjusted to the desired orientation and then secured in place. Postresection, the tracker is placed on the cut surface and the bone cut is measured by the navigation device (Fig. 2).

Fig. 1
figure 1

During use, the camera (A) detects movement of the trackers (B) within its field of view and relays information to the computer workstation (C), which sits outside of the sterile field. The workstation displays intraoperative data in real time and is controlled by the surgeon using buttons on the camera or by an assistant using the keyboard

Fig. 2
figure 2

Imageless computer navigation device. A With optical probe tracker slotted into the femoral cutting guide (w) changes in guide position are detected by the camera (not pictured) and the impact on planned cut parameters are displayed on the workstation (x), in real time. Insert shows that with extraarticular installation of the bone screw (y), the probe tracker can be placed on femur postresection (z) and bone cut parameters are displayed on the workstation in real time. B With optical probe tracker slotted into the tibial cutting guide (w) changes in guide position are detected by the camera (not pictured) and the impact on planned cut parameters are displayed on the workstation (x) in real time. Insert shows that with extraarticular installation of the bone screw (y), the probe tracker can be placed on tibia postresection (z) and bone cut parameters are displayed on the workstation in real time

The workflow for femoral cuts is similar to that for tibial cuts: a bone screw is drilled into the femur and the optical bone tracker is attached. Again, while the bone screw can be inserted either on the articular or extraarticular surfaces of the distal femur, cut verification requires extraarticular positioning of the screw. The femur is registered by rotating the hip joint through a range of motion to capture the hip center of rotation. The optical probe tracker is used to define the femur center, anterior–posterior femoral line, and the lateral and medial femoral condyles. As with the tibia, the optical probe tracker is attached to the cutting guide, providing real-time feedback to help position the cutting guide, which is secured with pins once the desired orientation is achieved. Postresection, the tracker is placed on the cut surface and the bone cut is measured by the navigation device (Fig. 2).

All resections were performed with an oscillating saw with a 1.27 mm saw blade. Multiple models of cutting guides were used; however, all were closed-slot guides designed for use with a 1.35 mm saw blade.

Outcome measures

Patient demographics at the time of surgery [age, sex, and body mass index (BMI)] were collected. Cut parameters were measured by the navigation device for the femur and tibia prior to and immediately following each cut, and included varus/valgus angle, flexion/extension, posterior slope, and resection depth. The primary analysis was a comparison of pinned cutting guide positions (planned resections, based on preoperative evaluation of individual patient deformity) to postresection measurements (actual resections) of the tibia and femur cut surface.

Statistical analysis

Data is expressed as mean ± standard deviation (SD). Paired Student’s t-tests were used to determine if there were any significant differences between mean pre- and postresection measurements. Alpha was set a priori at 0.05 for all statistical comparisons. All statistical analyses and calculations were performed using Microsoft Excel (Microsoft Corporation, Redmond, WA, USA).

Results

Study population

A total of 60 consecutive patients were identified for this study. Femur (n = 13) or tibia (n = 10) measurements were excluded from this study if the cutting guide or postresection cut measurement were not collected. Therefore, measurements of 47 femoral and 50 tibial resections were included in this study. No instances of loose cutting guides were reported.

Demographic data of patients included in the study is outlined in Table 1.

Table.1 Patient demographic data (N = 60)

Femoral measurements

For the femur, the mean planned varus/valgus cut based on cutting guide orientation was 0.32 ± 1.54°, while the mean actual cut was 0.04 ± 1.54° (p = 0.85, Table 2). The mean absolute difference between planned and actual cuts was 0.64 ± 0.52° in the varus/valgus plane (Table 3). The mean planned femoral flexion/extension cut was measured at 4.04 ± 1.84° of flexion, while the actual cut was measured at 3.45 ± 1.67° of flexion following resection (Table 2), with a mean absolute difference of 1.03 ± 0.96° (Table 3). The difference between planned and actual resection in the flexion/extension plane was statistically significant (p = 0.003). The mean planned medial and lateral resection cuts were 11.15 ± 2.97 mm and 9.61 ± 2.97 mm, respectively, while measurements following resection were 10.93 ± 2.74 mm (p = 0.32) and 9.18 ± 3.51 mm (p = 0.067, Table 2), with mean absolute differences of 1.05 ± 1.10 mm and 1.24 ± 1.04 mm, respectively (Table 3).

Table.2 Average cut parameters for femoral and tibial resection
Table.3 Average difference between planned and actual resection measurements for femoral and tibial resection

All differences between planned and actual resection in the varus/valgus plane were within 3°, while 94% (44/47) were within 3° for flexion/extension. Ninety-four percent (44/47) of measurements for medial and 96% (45/47) of lateral cutting guide deviations were within 3 mm (Fig. 3) [24]. The full range of data for these femoral measurements can also be seen in Fig. 2.

Fig. 3
figure 3

Distribution of deviation between planned and actual femoral resection in the A varus/valgus, B flexion/extension, C medial resection, and D lateral resection planes

Tibial measurements

For the tibia, the mean planned valgus/varus cut based on cutting guide orientation was 0.47 ± 1.26° varus, while the mean actual cut was 0.54 ± 1.23° (p = 0.63, Table 2). The mean absolute difference was 0.91 ± 0.79° in the varus/valgus plane (Table 3). Mean planned resection slope was 6.27 ± 2.07° versus a mean actual cut slope of 6.28 ± 2.63° following resection (p = 0.95, Table 2), with a mean absolute difference of 1.10 ± 1.00° (Table 3). Mean preresection measurements of planned medial and lateral resection cuts were 6.34 ± 2.95 mm and 8.53 ± 2.93 mm, respectively, while measurements following resection were 6.39 ± 2.12 mm (p = 0.78) and 8.59 ± 2.66 mm (p = 0.85, Table 2), with a mean average difference of 0.10 ± 1.79 mm and 0.16 ± 2.09 mm (Table 3). There were no significant differences between the planned resection parameters and actual resection.

Ninety-six percent (48/50) of differences between planned and actual resection in the varus/valgus plane were within 3°, and 96% (48/50) were within 3° for flexion/extension. Ninety-two percent (46/50) of measurements for medial and 90% (45/50) of lateral cutting guide deviations were within 3 mm (Fig. 4). The full range of data for these tibial measurements can also be seen in Fig. 3.

Fig. 4
figure 4

Distribution of deviation between planned and actual tibial resection in the A varus/valgus, B flexion/extension, C medial resection, and D lateral resection planes

Discussion

In primary TKA procedures, cutting errors during tibial and femoral resection have been shown to adversely impact component alignment [11,12,13]. Using an imageless computer navigation device, we quantified the extent to which planned and actual resection parameters differed, providing an indication of the magnitude of error associated with bone resection during primary TKA. We observed a statistically significant difference between planned and actual resection in the sagittal plane of the femur, and a difference between planned and achieved lateral femoral resection that trended toward significance. The data indicates that there is potentially important error associated with femoral and tibial cuts in TKA which may, in part, contribute to component and limb malalignment; additionally, this error can be assessed intraoperatively by imageless computer navigation.

We noted differences of 1.03° and 1.10° between planned and actual cuts in the femur and tibia, respectively, which mirror the results of similar studies. Chua et al. [25] noted errors of up to 1° in both tibial and femoral cuts, and determined these errors were potential causes of component malalignment noted on postoperative imaging. Other studies have identified alternative sources of component alignment error, including one study [26] that implicated cementing of components in place as a causal source of error of between 1° and 3°. Together, these findings suggest there are multiple sources of alignment error associated with resection cuts in TKA, which cumulatively push the error close to the threshold of ± 3° that is generally accepted as the maximal tolerance for postoperative alignment [9,10,11, 16,17,18]. Like other studies, we noted statistically significant differences in cuts in the femoral sagittal cuts [25]. Errors in the sagittal plane, especially in the femoral cut, have been reported to have an adverse impact on joint kinematics, range of motion, and component sizing, all of which adversely affect knee joint function [27]. Sagittal malpositioning of the femoral component has also been linked to inferior component survival and elevated risk for flexion contracture [7]. Tibial cut errors, meanwhile, threaten the accuracy of gap balancing and, therefore, the biomechanical stability of the new joint [28].

Drivers behind the greater frequency of sagittal plane cut errors are not fully known; however, one possibility is that sagittal plane resection may be more sensitive to differences in procedural factors. Plaskos et al. investigated these aspects, finding that the greater frequency of sagittal plane cut errors is influenced by surgeon experience, surgical instrumentation including cutting guide type, and cutting guide movement, all of which have been shown to impact the magnitude of resection error [11]. Deflection of the blade in the cutting guide represents another possible explanation for the observed differences. Several studies found that deflection or bending of the saw blade, especially when cutting through sclerotic bone, may lead to cutting error [29]. Poor bone mineral density in osteoporotic patients has also been suggested as a driver of poor tibial component position in CAS-assisted TKA, resulting from cutting guide pin motion during resection [30]. The use of slotted cutting guides or thicker blades have been suggested as potential solutions for these problems [11]; however, they do not completely eliminate the error associated with these cuts. In our study, we controlled for the hypothesized causal factors of surgeon experience, surgical instrumentation, and saw blade deflection by including only high-volume, experienced surgeons consistent in the use of 1.27 mm saw blades and 1.35 mm slotted cutting guides. However, we still observed differences between planned and actual cuts, suggesting that there are additional factors that contribute to the observed error that remain to be fully elucidated.

The use of computer-assisted navigation to confirm resection angles may provide an additional important cross check against inaccurate cuts. Previous studies have used navigation to monitor the orientation of cutting guides prior to actual cuts [25, 26], but the deflection that occurs during the cutting process warrants assessment of the actual cut angle itself prior to final component implantation. Further, the feedback provided by navigation devices may better characterize the impact that changes in surgical approach or equipment have on resection error, which could lead to alterations in the surgical workflow to help minimize cutting errors. For instance, the ability to verify and perform postresection adjustments may alter current approaches to preserving bone stock, which has been shown to decrease the risk of aseptic loosening, improve outcomes when the knee is in severe valgus [31], and increase the likelihood of procedural success in the event of revision surgery [32]. The option to modify initial cuts to achieve the desired plane of resection may lead surgeons to initially perform a more conservative resection to preserve bone stock, knowing that the feedback offered by navigation can help make postresection alterations if necessary. While this is one potential way in which resection verification and a better understanding of cutting errors could work to optimize TKA procedures, quantifying how cut verification alters resection procedure is outside the scope of this study; however, it but may be a potential avenue for future investigation.

The present study is not without limitations. Given the study design, we were unable to quantify what degree of cut variation may alter bone resection as a result of the real-time intraoperative feedback. Also, we did not evaluate component position on postoperative radiographs, which limited the ability to determine the final implant alignment, or to validate the feedback provided by the CAS. In this study, we sought only to quantify the magnitude of cutting errors that may occur during the procedure due to variables such as saw blade thickness, jig slot width, and technical variance. Additionally, although a power analysis shows that our sample size is akin to and consistent with similar studies, including Chua et al. and Plaskos et al., a larger sample size may improve the robustness and generalizability of our results [11, 15]. Future studies validating the feedback and assessing how this might alter the resections made during the TKA procedure would help better determine the clinical benefit of this feature.

Conclusion

Quantifying the difference between planned and actual resections in primary TKA procedures using an imageless navigation device demonstrated that small but significant deviations in resection occur. Unnoticed or uncorrected, these deviations may lead to component malalignment and deleterious effect on implant survivorship and patient outcomes. The use of intraoperative navigation offers the ability to verify cut angles to minimize resection errors, which may offer a viable path to optimizing component placement. Ultimately, improving component alignment by ensuring optimal cuts are achieved will improve long-term success of knee replacement and postoperative knee functional outcomes.

Availability of data and materials

Data sharing not applicable to this article as no datasets were generated or analysed during the current study.

References

  1. Feng JE, Novikov D, Anoushiravani AA, Schwarzkopf R (2018) Total knee arthroplasty: improving outcomes with a multidisciplinary approach. J Multidiscip Healthc 11:63–73

    Article  Google Scholar 

  2. Steinhaus ME, Christ AB, Cross MB (2017) Total knee arthroplasty for knee osteoarthritis: support for a foregone conclusion? HSS J 13(2):207–210

    Article  Google Scholar 

  3. Yau WP, Chiu KY (2008) Cutting errors in total knee replacement: assessment by computer assisted surgery. Knee Surg Sports Traumatol Arthrosc 16(7):670–673

    Article  CAS  Google Scholar 

  4. Longstaff LM, Sloan K, Stamp N, Scaddan M, Beaver R (2009) Good alignment after total knee arthroplasty leads to faster rehabilitation and better function. J Arthroplasty 24(4):570–578

    Article  Google Scholar 

  5. Choong PF, Dowsey MM, Stoney JD (2009) Does accurate anatomical alignment result in better function and quality of life? Comparing conventional and computer-assisted total knee arthroplasty. J Arthroplasty 24(4):560–569

    Article  Google Scholar 

  6. Stucinskas J, Robertsson O, Sirka A, Lebedev A, Wingstrand H, Tarasevicius S (2015) Moderate varus/valgus malalignment after total knee arthroplasty has little effect on knee function or muscle strength. Acta Orthop 86(6):728–733

    Article  Google Scholar 

  7. Kim YH, Park JW, Kim JS, Park SD (2014) The relationship between the survival of total knee arthroplasty and postoperative coronal, sagittal and rotational alignment of knee prosthesis. Int Orthop 38(2):379–385

    Article  Google Scholar 

  8. Sikorski JM (2008) Alignment in total knee replacement. J Bone Jt Surg Br 90(9):1121–1127

    Article  CAS  Google Scholar 

  9. Jeffery RS, Morris RW, Denham RA (1991) Coronal alignment after total knee replacement. J Bone Jt Surg Br 73(5):709–714

    Article  CAS  Google Scholar 

  10. Helmy N, Dao Trong ML, Kuhnel SP (2014) Accuracy of patient specific cutting blocks in total knee arthroplasty. Biomed Res Int 2014:562919

    PubMed  PubMed Central  Google Scholar 

  11. Plaskos C, Hodgson AJ, Inkpen K, McGraw RW (2002) Bone cutting errors in total knee arthroplasty. J Arthroplasty 17(6):698–705

    Article  Google Scholar 

  12. Otani T, Whiteside LA, White SE (1993) Cutting errors in preparation of femoral components in total knee arthroplasty. J Arthroplasty 8(5):503–510

    Article  CAS  Google Scholar 

  13. Nakahara H, Matsuda S, Moro-oka TA, Okazaki K, Tashiro Y, Iwamoto Y (2012) Cutting error of the distal femur in total knee arthroplasty by use of a navigation system. J Arthroplasty 27(6):1119–1122

    Article  Google Scholar 

  14. Mont MA, Urquhart MA, Hungerford DS, Krackow KA (1997) Intramedullary goniometer can improve alignment in knee arthroplasty surgery. J Arthroplasty 12(3):332–336

    Article  CAS  Google Scholar 

  15. Minns R (1992) Surgical instrument design for the accurate cutting of bone for implant fixation. Clin Mater 10(4):207–212

    Article  CAS  Google Scholar 

  16. Parisi TJ, Jennings JM, Dennis DA (2018) Coronal alignment in tka: traditional principles versus new concepts. Reconstr Rev. https://doi.org/10.15438/rr.8.2.213

    Article  Google Scholar 

  17. Bonner TJ, Eardley WG, Patterson P, Gregg PJ (2011) The effect of post-operative mechanical axis alignment on the survival of primary total knee replacements after a follow-up of 15 years. J Bone Jt Surg Br 93(9):1217–1222

    Article  CAS  Google Scholar 

  18. Cherian JJ, Kapadia BH, Banerjee S, Jauregui JJ, Issa K, Mont MA (2014) Mechanical, anatomical, and kinematic axis in TKA: concepts and practical applications. Curr Rev Musculoskelet Med 7(2):89–95

    Article  Google Scholar 

  19. Werner FW, Ayers DC, Maletsky LP, Rullkoetter PJ (2005) The effect of valgus/varus malalignment on load distribution in total knee replacements. J Biomech 38(2):349–355

    Article  Google Scholar 

  20. Elloy MA, Manning MP, Johnson R (1992) Accuracy of intramedullary alignment in total knee replacement. J Biomed Eng 14(5):363–370

    Article  CAS  Google Scholar 

  21. Toksvig-Larsen S, Ryd L (1994) Surface characteristics following tibial preparation during total knee arthroplasty. J Arthroplasty 9(1):63–66

    Article  CAS  Google Scholar 

  22. Foley KA, Muir JM. Improving accuracy in total knee arthroplasty: a cadaveric comparison of a new surgical navigation tool, Intellijoint KNEE, with computed tomography imaging. Intellijoint Surgical, Inc., Waterloo, ON, 2019

  23. Hsu RW, Hsu WH, Shen WJ, Hsu WB, Chang SH (2019) Comparison of computer-assisted navigation and conventional instrumentation for bilateral total knee arthroplasty: the outcomes at mid-term follow-up. Medicine (Baltimore) 98(47):e18083

    Article  Google Scholar 

  24. Winemaker MJ (2002) Perfect balance in total knee arthroplasty: the elusive compromise. J Arthroplasty 17(1):2–10

    Article  CAS  Google Scholar 

  25. Chua KH, Chen Y, Lingaraj K (2014) Navigated total knee arthroplasty: is it error-free? Knee Surg Sports Traumatol Arthrosc 22(3):643–649

    Article  Google Scholar 

  26. Catani F, Biasca N, Ensini A, Leardini A, Bianchi L, Digennaro V et al (2008) Alignment deviation between bone resection and final implant positioning in computer-navigated total knee arthroplasty. J Bone Jt Surg Am 90(4):765–771

    Article  Google Scholar 

  27. Koh YG, Lee JA, Lee HY, Suh DS, Kim HJ, Kang KT (2019) Effect of sagittal femoral component alignment on biomechanics after mobile-bearing total knee arthroplasty. J Orthop Surg Res 14(1):400

    Article  Google Scholar 

  28. Brooks P (2009) Seven cuts to the perfect total knee. Orthopedics. https://doi.org/10.3928/01477447-20090728-27

    Article  PubMed  Google Scholar 

  29. Laskin RS, Beksac B (2006) Computer-assisted navigation in TKA: where we are and where we are going. Clin Orthop Relat Res 452:127–131

    Article  Google Scholar 

  30. Lee D, Padhy D, Lee S, Nha K, Park J, Han S (2012) Osteoporosis affects component positioning in computer navigation-assisted total knee arthroplasty. Knee 19(3):203–207

    Article  Google Scholar 

  31. Schnurr C, Csecsei G, Nessler J, Eysel P, Konig DP (2011) How much tibial resection is required in total knee arthroplasty? Int Orthop 35(7):989–994

    Article  Google Scholar 

  32. Haas SB, Nelson CL, Laskin RS (2000) Posterior stabilized knee arthroplasty: an assessment of bone resection. Knee 7(1):25–29

    Article  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

Not applicable.

Author information

Authors and Affiliations

Authors

Contributions

R.S., J.M.M., and I.R.L. jointly designed this study. R.S., M.M., and S.E.M. generated and collected all data. I.R.L. and J.M.M. performed all analyses. M.A.Z., I.R.L., J.M.M., and R.S. prepared the manuscript. All authors reviewed, read and approved the final manuscript.

Corresponding author

Correspondence to Ran Schwarzkopf.

Ethics declarations

Ethics approval and consent to participate

The study received IRB approval from the institutional ethics board prior to data collection.

Consent for publication

Not applicable.

Competing interests

I.R.L. and J.M.M. are employees of Intellijoint Surgical, Inc. and hold stock or stock options. R.S. has received consultancy fees from Intellijoint Surgical, Inc. and holds stock or stock options. M.M. has received consultancy fees from Intellijoint Surgical, Inc. S.E.M. has received consultancy fees from Intellijoint Surgical, Inc. and has been a paid speaker or presenter.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Schwarzkopf, R., Meftah, M., Marwin, S.E. et al. The use of imageless navigation to quantify cutting error in total knee arthroplasty. Knee Surg & Relat Res 33, 43 (2021). https://doi.org/10.1186/s43019-021-00125-z

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s43019-021-00125-z

Keywords