Skip to main content

Table 4 Comparison of gap size and inclination without the distal femoral trial component versus with the final femoral trial component

From: Evaluation of the flexion gap with a distal femoral trial component in posterior-stabilized total knee arthroplasty

 

Without distal femoral trial component

Final femoral trial component

P value

Joint gap size (mm)

16.7 ± 0.45

16.1 ± 0.44

0.033

Joint gap inclination (°)

5.4 ± 0.46

4.6 ± 0.4

0.016